Contact

info@airsensepro.com

— Measurement Performance Guide

Breathalyzer Accuracy and Repeatability

How to Evaluate Measurement Performance in Real-World Testing

Accuracy tells you how close a breathalyzer reading is to a known reference. Repeatability tells you how consistently the instrument produces similar results when the same test is repeated under the same conditions.

Breathalyzer Accuracy and Repeatability

— Two different concepts

Accuracy vs Repeatability

A device can produce very similar readings every time and still be wrong. Understanding both concepts is essential when evaluating a breathalyzer.

02-accuracy-target

01 / CLOSE TO THE

REFERENCE

Accuracy

How close is the reading to the reference?

Accuracy describes how closely a measured value agrees with a known reference value under the specified test conditions.


Reference: 0.050

0.050

0.051

0.050

0.049

0.050

02 CLOSE TO EACH OTHER

02 / CLOSE TO EACH

OTHER

Repeatability

How consistent are repeated measurements?

Repeatability describes how closely repeated readings agree with each other when the same procedure is repeated under the same conditions.


Same Test Conditions

0.060

0.061

0.060

0.061

0.060

Accuracy and Repeatability Are Not the Same

Fuel Cell Platforms

Fuel cell sensing systems do not rely on maintaining a heated semiconductor surface, so many designs can begin testing with little or no conventional warm-up.

Test 1 0.050
Test 2 0.051
Test 3 0.050
Test 4 0.049
Test 5 0.050

Good Accuracy · Good Repeatability

Semiconductor Platform

Reference: 0.050

Test 1 0.060
Test 2 0.061
Test 3 0.060
Test 4 0.061
Test 5 0.060

Good Repeatability · Poor Accuracy

Illustrative example only. Acceptance limits should always be evaluated against the applicable product specification or test standard.

Practical test guide

How to Test Breathalyzer Accuracy

A meaningful accuracy check requires a known alcohol reference and controlled test conditions. An unknown human breath sample cannot tell you whether the instrument itself is accurate.

Use the procedure specified for the breathalyzer model and the applicable test standard.

01
Prepare a Known Reference

Use an appropriate certified alcohol reference such as a dry gas standard or wet bath simulator according to the product procedure.

02
Control Test Conditions

Confirm device condition, calibration status, power level, sampling path and environmental conditions before testing.

03
Run the Test

Introduce the known reference using the specified test or calibration method and allow the instrument to complete the measurement cycle.

04
Record the Reading

Document the displayed value and any relevant conditions so repeated tests can be compared consistently.

05
Compare With the Reference

Evaluate the measured value against the known reference and the tolerance specified for the model or test protocol.

Test Breathalyzer Accuracy

Known Alcohol Reference

Stable Test Conditions

Breathalyzer Test

Record Reading

Compare With Reference

Repeatability testing

How to Evaluate Repeatability

Repeatability requires multiple measurements performed using the same procedure under consistent conditions. One reading can suggest agreement with a reference, but it cannot show how consistent the instrument is.

Good Repeatability

Good Repeatability

Repeated readings remain closely grouped when the same test is repeated under controlled conditions.

Reference: 0.050

0.050

0.051

0.050

0.049

0.050

Poor Repeatability

Readings vary widely even though the same test procedure and reference are used.

Reference: 0.050

0.043

0.057

0.049

0.061

0.045

Repeatability testing

How to Interpret Measurement Performance

Accuracy and repeatability answer different questions, so they should be evaluated separately.

Accuracy

Compare the measured value with the known reference value.

Measurement Error = Measured Value − Reference Value

Example
Measured value: 0.052
Reference value: 0.050
Measurement error: +0.002

Whether the error is acceptable depends on the tolerance specified for the product model or applicable test standard.

Repeatability

Compare repeated readings with each other under the same test conditions.

Smaller Variation Between Repeated Readings = Better Repeatability

Example
0.050 · 0.051 · 0.050 · 0.049 · 0.050

Closely grouped readings indicate better repeatability than widely scattered results.

Advanced evaluation may use statistical measures such as range or standard deviation depending on the test protocol.

What changes the result?

Factors That Affect Accuracy & Repeatability

Measurement performance depends on the entire test system—not only the nominal specification of the alcohol sensor.

Breath Sample

Breath volume, flow and sampling quality can influence the amount of alcohol reaching the sensor.

Calibration Status

A device that is overdue for calibration may drift away from its intended measurement response.

Sensor Condition

Sensor age, exposure history and recovery behavior can influence performance over time.

Test Interval

Repeated tests performed too quickly may not allow the sensing system sufficient recovery time.

Temperature

Operating outside the intended temperature range can affect sensor and electronic behavior.

Sampling System

Pumps, airflow paths and mouthpiece installation can affect how the sample reaches the sensor.

Electronics & Algorithm

Signal amplification, filtering and data conversion influence the final displayed reading.

Environmental Interference

Some vapors and environmental conditions may affect measurement depending on the sensor technology.

A common misunderstanding

Human Breath vs Reference Standard Testing

Human breath is useful for practical use testing, but it is not a controlled reference for evaluating instrument accuracy.

Human Breath Testing

A real breath sample can vary from one test to the next because of multiple physiological and procedural factors.

  • Breathing pattern
  • Breath volume and flow
  • Recent alcohol consumption
  • Residual mouth alcohol
  • Time between tests
  • Normal biological variation

Reference Standard Testing

A known reference provides a controlled basis for checking whether the instrument is measuring within its intended range.
  • Breathing pattern
  • Defined test procedure
  • Controlled conditions
  • Repeatable test input
  • Suitable for accuracy evaluation
  • Useful for calibration verification

You cannot accurately evaluate a breathalyzer using an unknown breath sample.

Avoid false conclusions

Common Breathalyzer Testing Mistakes

Many apparent “accuracy problems” come from uncontrolled testing rather than from the instrument itself.

Testing Immediately After Drinking

Residual alcohol in the mouth can temporarily produce an unusually high reading.

Using a Person as the Reference

Human breath alcohol is not a fixed known concentration and cannot act as a calibration reference.

Repeating Tests Too Quickly

The sensor and sampling system may require recovery time between repeated measurements.

Comparing Two Devices Without a Standard

If two devices disagree, a known reference is needed to determine which reading is closer to the target.

Ignoring Calibration Status

A comparison is less meaningful if one or both instruments are overdue for calibration.

Changing Test Conditions

Temperature, airflow, sampling technique and other conditions should remain as consistent as possible.

Airsense quality perspective

How Airsense Evaluates Measurement Performance

Airsense approaches breathalyzer performance as a system-level result involving the sensor, sampling path, electronics, calibration and measurement logic.

For formal production and quality-control claims, the exact test procedure should be based on the applicable Airsense model specification and internal quality documentation.

Reference Testing
Known reference input for measurement checks.

Repeat Measurement

Multiple tests to evaluate consistency.

Calibration Verification
Confirming measurement response against the intended range.

Production QC Known reference input for measurement checks.
How to Test Breathalyzer Accuracy

Different applications

Performance Requirements Depend on How the Breathalyzer Is Used

A personal breathalyzer and a high-frequency professional screening device do not necessarily need the same architecture or service strategy.

Personal Testing

Compact design, convenience and appropriate measurement performance for individual use.


Airsense examples: K1 / H3 / A5

Professional Screening

Stronger emphasis on repeatability, sampling control and reliable operation across repeated tests.


Airsense example: H3 Pro

High-Frequency Professional Use

Greater focus on data handling, maintainability, sensor service and repeated daily testing.


Airsense example: S1

Frequently asked questions

Accuracy & Repeatability FAQ

Common questions about evaluating breathalyzer measurement performance.

Why do two breathalyzer readings sometimes differ?

Differences can result from sampling technique, test interval, sensor recovery, calibration status, environmental conditions or normal variation in the breath sample.

Not reliably. If two devices disagree, a known alcohol reference is needed to determine which instrument is closer to the target value.

No. A device can repeatedly produce almost identical readings that are consistently above or below the true reference value.

The correct number depends on the product specification or evaluation protocol. Repeated measurements should follow a defined procedure rather than an arbitrary number of uncontrolled human breath tests.

No. Fuel cell technology provides advantages in ethanol selectivity and stability, but complete measurement performance also depends on sampling, electronics, calibration, algorithms and maintenance.

Calibration intervals vary by model, usage frequency and application. Follow the schedule specified in the product manual or provided by Airsense technical support.

Need Help Evaluating Breathalyzer Performance?

ell us your application, testing frequency and measurement requirements. Airsense can help you select an appropriate breathalyzer platform and explain the relevant test and calibration approach.